Options (AMXR)
Switch TypesSelectable for each switch channel separately
Standard
Dual-level switch
- Output switches between two voltage levels Vneg and Vpos
- Vneg ≤ Vpos must be fulfilled at all times
Option T (Trilevel)
Three-level switch
- Output switches between three voltage levels Vneg, Vmid, and Vpos
- Vneg ≤ Vmid ≤ Vpos must be fulfilled at all times
- For applications such as injection, compression and ejection of ions
IsolationSelectable for each switch channel separately
Standard
- Maximum isolation voltage: 2.1 kV
- Maximum burst frequency: > 10 MHz
- Minimum pulse duration: < 50 ns
- Propagation delay: 65 ns
- Jitter: < 1 ns
Option H (HV Isolation)
High-voltage isolation
- Maximum isolation voltage: 3 kV (5 kV or more upon request)
- Maximum burst frequency: 2 MHz
- Propagation delay: 120 ns
Power SupplyModification of entire device
Standard
- Nominal operating frequency of the auxiliary switching regulators: 125 kHz
- Optional dithering control
Option N (Low Noise)
Power supply optimized for low noise level at the switch output
- Shielded transformers
- Nominal operating frequency of the auxiliary switching regulators: about 28 kHz
- Optional dithering control
Switch ControlModification of entire device
Standard
Switch input: BNC socket on front panel
Option L (LEMO Connector)
Switch input: coaxial LEMO socket on the front panel
Option R (Rear Control)
Switch control via connector on the rear panel
Switch inputs (selectable via a slide switch):
- BNC/LEMO socket on the front panel
- Control signal at the rear connector
Option M (Monitoring)
Only available as Option RM:
Rear control (option R) + monitoring
Rear control (option R) + monitoring
BNC/LEMO socket on the front panel can be used as a monitoring output (selectable via a slide switch); in this case, the signal from the rear connector must be used for controlling the switch